medical Class II Updated 2018-11-28

Voluntis recalls Insulia Diabetes Management Companion

Recalled Product

Insulia Diabetes Management Companion


Hazard / Issue

The firm identified an issue with the basal calculator identified on the Android version of the Insulia application. The bug could lead from low-impact to high-impact hypoglycemia depending on the circumstances. A new update of Insulia is now available (version 1.6.42), which resolves this issue.

Issued by

FDA

Distribution: Distributed to accounts in Arizona, Masachusetts, Michigan, Ohio, Tennessee, and Vermont. International distribution to Canada and France (however, French customers are not affected by this recall).
Lot/Code Info: Software version numbers 1.4.3 / 1.6.3 / 1.6.31 / 1.6.40