medical Class II Updated 2018-11-28

Ascom (US) Inc. recalls Ascom Telligence Nurse Call System

Recalled Product

Ascom Telligence Nurse Call System


Hazard / Issue

There is an issue that could cause the Telligence System Gateway II, NGGTWY2 H, to stop communicating to the hallway devices connected to it. When this occurs, the dome lights connected to the Gateway II Ports A and B will start flashing, indicating that they are no longer communicating to the Gateway II. If this occurs, device failures that WOULD otherwise be annunciated WILL NOT be indicated at staff consoles that are configured to display failures.

Issued by

FDA

Affected States: AL, CA, CO, GA, IN, KY, LA, ME, MI, MO, MN, MS, NE, NV, OH, PA, RI, TX, UT, VA, WI
Lot/Code Info: Version 5.0 (Build 5.0.4)
View official government recall