medical Class II Updated 2023-12-13

Stanbio Laboratory, LP recalls STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB M

Recalled Product

STAT-Site M BHB Test Strips, used with STAT-SITE M Beta-HB Meter as an in vitro diagnostic for blood ketone values.


Hazard / Issue

Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00657498000090, Lot/Expiration: 203044/2024-01-31, 209644/2024-06-30