medical Class II Updated 2018-11-28

Becton Dickinson & Company recalls KIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine

Recalled Product

KIT URIN CUP PLC 16X100 10.0 UA YEL, Cat. No. 364981: Urine Collection Kit for collection and transport of urine specimens for urinalysis


Hazard / Issue

BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.

Issued by

FDA

Distribution: Illinois
Lot/Code Info: All lots within expiry