Becton Dickinson & Company recalls KIT URIN CUP 16X100 8.0 CONI UA YEL, Cat. No. 364989 BD Vac
Recalled Product
KIT URIN CUP 16X100 8.0 CONI UA YEL, Cat. No. 364989 BD Vacutainer Urine Cup Kit with UA Tube for Midstream Specimens
Hazard / Issue
BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
Issued by
FDA
Were you affected by this recall?
If you were harmed by a recalled product, you may have legal options. Consider speaking with a consumer protection lawyer in your state.
Looking for safe alternatives?
Browse certified-safe products in this category on Amazon →
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.
CAN-SPAM compliant · No spam · Unsubscribe anytime