medical Class II Updated 2023-12-13

Ortho-Clinical Diagnostics, INc. recalls VITROS Chemistry Products ECO2 Slides-For in vitro diagnosti

Recalled Product

VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides quantitatively measure total carbon dioxide (CO2) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600 XT 3400 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. Product Code: 8262396


Hazard / Issue

Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 10758750004645 Lot Numbers/Exp. Date: 5732-3598-4649 01-Feb-2025; 5732-3598-4650 01-Feb-2025; 5732-3598-4651 01-Feb-2025; 5732-3598-5102 01-Mar-2025; 5732-3598-5105 01-Mar-2025; 5732-3598-5189 01-Mar-2025
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →