EUROTROL INC recalls HemoTrol Level 2- In Vitro Diagnostic Hematology quality con
Recalled Product
HemoTrol Level 2- In Vitro Diagnostic Hematology quality control mixture Catalogue number: 171.002.002 Eurotrol HemoLin is an assayed hemoglobin control intended for professional use in the verification of the linearity of the HemoCue¿ Hb 201 systems
Hazard / Issue
Incorrect measurement results caused by microbial contamination of the product, results in values lower than assigned values