medical Class II Updated 2020-12-09

Invacare Corporation recalls Invacare PerfectO2 V Oxygen Concentrator- intended for patie

Recalled Product

Invacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5


Hazard / Issue

Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model numbers: IRC5PO2V from Serial # 14HF039890 to 18JF007415; IRC5PO2VC from Serial # 14IF004028 to 18JF028633; IRC5PO2VAW from serial #17HF030338-18IF018522; IRC5PO2VAWL2 from serial # 18BF007341-18IF024451; IRC5PO2VAWL3 from serial # 17JF002563-19AF008559; IRC5PO2VAWL5 from serial # 18EF003466-18EF003544. DI Number: 00841447100249