medical Class II Updated 2022-02-02

Siemens Medical Solutions USA, Inc recalls Artis pheno with software VE20C- diagnostic imaging angiogr

Recalled Product

Artis pheno with software VE20C- diagnostic imaging angiography system Model: 10849000


Hazard / Issue

Four potential software issues with Artis pheno and Artis icono systems with software VE20C: 1. Updated calibration data not saved with measurement after scene+/-; 2. No x-ray possible, system shutdown/restart might be required during intervention; 3. Corrupted Image during Roadmap; 4. Unintended shutdown of Imaging System with UPS (Uninterruptable Power Supply) option

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 164719 164712 164722 164723 164728 164711 164718 164703 164721 164726 164724 164725 UDI: 04056869046877