medical Class II Updated 2023-12-20

Meridian Bioscience Inc recalls Para-Pak Clean Vial-For the collection, transportation, pres

Recalled Product

Para-Pak Clean Vial-For the collection, transportation, preservation, and examination of stool specimens containing intestinal parasites Catalog Number: 900312


Hazard / Issue

Plastic vials contain a manufacturing defect may cause an increased risk of leakage and presents a risk of incidental exposure to biological hazards associated with direct contact with the patient specimen.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 00840733101892 Lot Numbers: 509940Q Expiration Date: 09/12/2030