medical Class II Updated 2024-12-04

Smiths Medical ASD, Inc. recalls smiths medical PORTEX Tracheal Tube Exchange Guide, 10Ch x

Recalled Product

smiths medical PORTEX Tracheal Tube Exchange Guide, 10Ch x 70cm, REF 14-504-76


Hazard / Issue

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: UDI/DI 35021239002593, Lot Numbers: 0003373, 0003456, 0003621, 0003374, 0003376, 0003375, 0003459, 0003391, 0003480, 0003549, 0003301, 0003124, 0003130, 0003139, 0003140, 0003144, 0003148, 0003152, 0003161, 0003169, 0003185, 0003205, 0003210, 0003222, 0003232, 0003236, 0003255, 0003276, 0003277, 0003315, 0003329, 0003339, 0003367, 0003370, 0003371, 0003372, 0003381, 0003395, 0003399, 0003403, 0003414, 0003449, 0003457, 0003469, 0003476, 0003477, 0003489, 0003492, 0003500, 0003573, 0003484, 0003619, 0003557, 0003510, 0003522, 0003597, 0003511, 0003622, 0003553, 0003581, 0003605, 0003529,
View official government recall