medical Class II Updated 2022-02-09

LivaNova USA, Inc. recalls Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, M

Recalled Product

Vagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mm electrode inner diameter) and Model 303-30 PereniaDURA Lead (3.0 mm electrode inner diameter)


Hazard / Issue

A manufacturing issue may cause the silicone tubing wall thickness of vagus nerve stimulation leads to be less than the design specification. If the tubing becomes completely breached, the patient may experience discomfort or pain due to delivery of stimulation to an unintended location. If the inner lead conductor is exposed to body fluids the lead may break causing loss of therapy.

Issued by

FDA

Affected States: NE, MO, AR, WV, KS, MA, IN
Lot/Code Info: UDI: 05425025750115. Serial Numbers: 119441, 119442, 119443, 119452, 119453, 119454, 119455, 119456, 119457, 119458
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →