medical HIGH RISK Updated 2023-12-27

ResMed Ltd. recalls AirFit F30i Full Face Mask and User Guide

See all recalls from ResMed Ltd. →

Recalled Product

AirFit F30i Full Face Mask and User Guide

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All mask lots used with User Guide 638225/version 2020-02, and prior. UDI-DI/GTIN: 619498633326, 619498633067, 619498633012, 619498633340, 619498633302, 619498633333, 619498633050, 619498633319, 619498633005, 619498633135, 619498633128, 619498633364, 619498633388, 619498633371, 619498633401, 619498633395, 619498633104, 619498633142, 619498633111, 619498633081, 619498633074

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for AirFit F30i Full Face →
View official government recall