Shimadzu Medical Systems recalls Trinias, Digital Angiography System containing a MH-200, Cei

Recalled Product

Trinias, Digital Angiography System containing a MH-200, Ceiling Suspended C-arm Support - Product Usage: intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.


Hazard / Issue

The firm has identified a problem with the control software for the celling arm of the X-Ray System . This software issue could cause the ceiling arm of the device to move after the operator has released the movement lever. This could result in possible collision or harm to the patient or system operator.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Affected Serial Numbers: 41EFCD7A6001 41EFC73A5001, 41EFCD79C001, 41EFD3B9C001, 41EFFF79C001, 41EFD3B9A001, 41EFCD79A001, 41EFFF79A001, 41F005B91001, 41F005B8C001, 41EFCD78B001, 41EFCD77C001, 41E58C776001, 41E58C777001, 41E58C775001, 41E58C773001, 41E58C76B001, 41E58C769001, 41E58C766001, 41E58C765001, 41E58C757001, 41B24774C001, 41B247746001,