medical Class II Updated 2023-12-27

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Allura systems with monoplane fixed ceiling mounts:

Recalled Product

Philips Allura systems with monoplane fixed ceiling mounts: 722001 Allura Xper FD10C 722003 Allura Xper FD10 722006 Allura Xper FD20 722010 Allura Xper FD10 722012 Allura Xper FD20 722015 Allura Xper FD20 OR Table 722016 Integris H5000 C / Allura 9C 722018 Integris Allura 9 722022 Allura Xper FD10 OR Table 722023 Allura Xper FD20 OR Table 722026 Allura Xper FD10 722028 Allura Xper FD20 722033 Allura Xper FD10 OR Table 722035 Allura Xper FD20 OR Table 722043 Integris Allura 15 & 12 (monoplane) 72246 Poly G - OMCP - VISUB - CCD (H5000)


Hazard / Issue

Ceiling mounted L-arm contains a rotation cover that may potentially be susceptible to falling if a collision between the L-arm and other hospital equipment (i.e., an operating light) were to occur result in injury or potential sterility issues due to the cover becoming loose

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model/Name/UDI-DI: 722001 Allura Xper FD10C; 722003 Allura Xper FD10; 722006 Allura Xper FD20 (01)00884838059115(21); 722010 Allura Xper FD10 (01)00884838059030(21); 722012 Allura Xper FD20 (01)00884838059054(21); 722015 Allura Xper FD20 OR Table; 722016 Integris H5000 C / Allura 9C ; 722018 Integris Allura 9; - 722022 Allura Xper FD10 OR Table; 722023 Allura Xper FD20 OR Table (01)00884838059085(21;) 722026 Allura Xper FD10 (01)00884838054189(21); 722028 Allura Xper FD20 (01)00884838054202(21); 722035 Allura Xper FD20 OR Table (01)00884838054240(21); 722043 Integris Allura 15 & 12 (monoplane).
View official government recall