medical Class II Updated 2024-12-04

LivaNova Deutschland GmbH recalls Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Softwar

Recalled Product

Essenz HLM (Heart-Lung Machine), REF: 49-00-10, with Software


Hazard / Issue

Heart-Lung Machine GUI touchscreen may go dark for 10-15 seconds, safety/performance functions remain operational before reset, control maintained through backup control panel, but second reset may occur, may lead to gas blender standby, then gas blender must be reactivated using user interface on blender, sensors deactivated until new case started; GUI reset can lead to Hypoxia and Hypoperfusion.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 04033817903062. Software Version: 1.5. Serial Numbers: 49PK00436, 49PK00108, 49PK00374, 49PK00479, 49PK00480, 49PK00481, 49PK00093, 49PK00091, 49PK00092, 49PK00163, 49PK00497, 49PK00498, 49PK00219, 49PK00223, 49PK00224, 49PK00225, 49PK00226, 49PK00241, 49PK00255, 49PK00292, 49PK00291, 49PK00293, 49PK00490, 49PK00487, 49PK00488, 49PK00278, 49PK00274, 49PK00277, 49PK00275, 49PK00276, 49PK00273, 49PK00457, 49PK00458, 49PK00459, 49PK00455, 49PK00456, 49PK00290, 49PK00288
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →