medical Class II Updated 2023-12-27

Karl Storz Endoscopy recalls Injection Needle, LUER-lock,0.8mm, REF: 26175PD

Recalled Product

Injection Needle, LUER-lock,0.8mm, REF: 26175PD


Hazard / Issue

The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.

Issued by

FDA

Affected States: MI, NH, SC, PA, FL, NY, NC, MS, MO, CA, NJ, IL, NV, OK, CO, OH, WA, WI, VA, IN, IA, AR, TX, WV, MD, MN, GA, MA, NM, VT, AK, OR, ID, LA, KS, DC, KY, HI
Lot/Code Info: UDI-DI: 04048551099391. All lot numbers.
View official government recall