medical Class II Updated 2019-12-04

Abbott GmBH & Co. KG recalls Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott b

Recalled Product

Alinity c Carbon Dioxide Reagent Kit, Produced for Abbott by Sekisui Diagnostics P.E.I. Inc. Prince Edward Island Canada


Hazard / Issue

Abbott internal studies have determined that atmospheric Carbon Dioxide (CO2) can be absorbed into the Alinity c Carbon Dioxide reagent, resulting in the potential for incorrect results. Abbott was made aware of this issue from customer complaints. Internal testing demonstrated that the amount of CO2 absorbed is higher with increased reagent carousel rotation and when the volume of reagent in the cartridge is reduced. This phenomenon can be detected as a shift in QC. The shift is predictive in both direction and magnitude of the potential impact to patient results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: REF # 07P7220 Lot # 54454UQ03 and REF # 07P7230 Lot # 54449UQ03
View official government recall