medical Class I Updated 2020-12-30

Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL1-HC

Recalled Product

Flexor Check-Flo Introducer, RPN KCFW-4.0-35-55-RB-HFANL1-HC, Order Number G56234 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.


Hazard / Issue

There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 13047588, 13050227, 13050229, 13130291, 13138925, 13203167, 13226780, 13276230, 13315165, 13332154, 13375204, 13375207