medical Class II Updated 2019-12-11

NDDd Medical Technologies, Inc. recalls EasyOne Pro Respiratory Analysis System Pulmonary Function T

Recalled Product

EasyOne Pro Respiratory Analysis System Pulmonary Function Testing Device, Model 3000-1


Hazard / Issue

The CO/CO2 sensor cable harness may have lower electromagnetic shielding, which might have resulted in a non-compliance with the IEC 60601-1-2 standard. As a consequence of a lower electromagnetic shielding provided by the cable harness, the CO/CO2 sensor could have been influenced by high frequencies emitted by very close devices.

Issued by

FDA

Affected States: KY, MO, NY, OH, PA
Lot/Code Info: Serial Numbers: 550066 550067 550068 550069 550071 550072
View official government recall