medical Class I Updated 2020-12-30

Cook Inc. recalls Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL1-HC

Recalled Product

Flexor Check-Flo Introducer, RPN KCFW-8.0-35-70-RB-HFANL1-HC, Order Number G56228 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.


Hazard / Issue

There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Numbers: 13229667, 13231805, 13254613, 13291858, 13291869, 13310725, 13318020, 13337787, 13337798, 13346070, 13354237, 13366209, 13369431, 13375208, 13391162, 13407499, 13435157, NS13401856,