medical Class II Updated 2020-12-23

Carl Zeiss Meditec AG recalls Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Us

Recalled Product

Carl Zeiss Meditec Ag VISULENS 550 REF 2268-038 - Product Usage: is a device for measuring the refractive power of contact lenses and spectacle lenses and displaying measurement values for sphere, cylinder and prism.


Hazard / Issue

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending
View official government recall