Abbott Laboratories recalls Alinity ci-series System Control Module, REF: 03R70-01

Recalled Product

Alinity ci-series System Control Module, REF: 03R70-01


Hazard / Issue

There are potential performance issues found in the Alinity ci-series System software versions 3.4.0 and lower, including: 1) SCC Reagent load error during daily maintenance; 2) Sample Laboratory Report error; 3) Customer requests mechanism to prevent user error when creating new lots of value assigned Calibrators; 4) Error in processing tests after CCCintController encountered an error with the assay information downloaded from SCC; 5) The assay editor does not correctly check sample dilution total volume limits; 6) User-Applied Labels not adhering to Reagent Bottles; 7) ICT Module did not expire after warranty was exceeded. These performance issues could lead to erroneous results for multiple analytes.

Issued by

FDA

Affected States: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV
Lot/Code Info: UDI-DI: 00380740137410. Software Version 3.4.0 and lower.