medical Class II Updated 2020-12-23

Carl Zeiss Meditec AG recalls Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usag

Recalled Product

Carl Zeiss Meditec AG VISUREF150 REF 2227-967 - Product Usage: is a diagnostic instrument used to determine the objective refractive values and keratometric characteristics of the human eye. The results are used to assist the process of prescribing optical aids such as eyeglasses and contact lenses.


Hazard / Issue

Due to the incorrect DC/DC converter being assembled onto the communication board. As a result the effectiveness of the communication board cannot be guaranteed in that the safety function of the board would be insufficient resulting in an increased leakage of current and potentially cause electrical shocks to the user.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model: VISUREF 150 Catalog Number: 000000-2227-967 Serial Numbers: Pending