medical Class II Updated 2024-01-03

Atrium Medical Corporation recalls Atrium Oasis Drain Single W / AC, Part No. 3600-100

Recalled Product

Atrium Oasis Drain Single W / AC, Part No. 3600-100


Hazard / Issue

A sterilization nonconformance prematurely aged the product by 48.3 days, resulting in the expiration date on the label not being accurate. This nonconforming product was inadvertently distributed to customers. Increased risk to patients and healthcare providers using the product after the shortened expiration date of November 17, 2025.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI 00650862110012 Lots 492065 and 492075