medical Class II Updated 2022-02-16

Synthes (USA) Products LLC recalls Universal Chuck-The Universal Chuck is an Instrument Handle

Recalled Product

Universal Chuck-The Universal Chuck is an Instrument Handle provided as an alternative instrument available for use in the TN-ADVANCED" Tibial Nailing System (TNA), TFN-ADVANCED" Proximal Femoral Nailing System (TFNA), RFN-ADVANCED" Retrograde Femoral Nailing System (RFNA), FRN-ADVANCED" Femoral Recon Nailing System (FRNA), and Flexible Monobloc Reamers. Instrument Handles are intended for holding instruments for manipulation during orthopedic surgery. Product No.: 03.043.001


Hazard / Issue

Top cap may loosen and detach if the Universal Chuck becomes jammed and the user attempts to manually free the device, internal ball bearings may fall out onto the surgical field and cause surgical delay or, and if not recovered, adverse tissue reaction and infection.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot/Serial No.: 20013101, 20052701, 20228901, 20229301, 20253901, 20558602, 20698102, 20777201 21869701, 21983904, 21995301 GTIN: 10886982297372
View official government recall