medical Class II Updated 2018-12-26

Advanced Instruments, LLC recalls GloCyte RBC Reagent - Product Usage: The GloCyte Automated C

Recalled Product

GloCyte RBC Reagent - Product Usage: The GloCyte Automated Cell Counter System is intended for use by trained healthcare professionals in clinical laboratories to provide quantitative determination of fluorescence labeled total nucleated cells and erythrocytes in cerebrospinal fluid collected from adult and pediatric patients. The GloCyte Low and High Level Controls are assayed hematology controls designed to monitor the performance of the GloCyte Automated Cell Counter System. Assayed parameters include total nucleated cells and erythrocytes.


Hazard / Issue

RBC Reagent failure. High fluorescent counts during RBC Reagent + Diluent QC check that will result in failed QC check, preventing user from testing patient samples for RBC.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Part number 111011, Lot # R1513
View official government recall