medical Class II Updated 2020-12-30

CryoLife, Inc. recalls CryoValve SG Pulmonary Human Heart Valve. The valve is cryo

Recalled Product

CryoValve SG Pulmonary Human Heart Valve. The valve is cryopreserved in a tissue culture medium containing cryoprotectants within the innermost pouch of a three-pouch packaging system. The packaging system not only withstands ultracold temperatures, but also allows for aseptic introduction of the valve into the operating environment. Supercooling by liquid nitrogen boost is begun prior to crystallization to minimize ice crystal damage to the valve matrix. Finally, the valve is transferred to a liquid nitrogen freezer for long-term storage at or below 135¿C. CryoValve SG is distributed in two anatomic configurations: pulmonary valve and conduit, and pulmonary valve.


Hazard / Issue

On November 10, 2020, it was discovered that there are post- processing microbial culture test failures for other allografts, processed using a TRIS saline solution lot that was also used in processing allografts subjected to the recall.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model/Catalog Number: SGPV00 Serial Numbers: 11685214, 11685554, and 11682491 UDI: 00877234000447
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →