medical Class II Updated 2022-03-02

BALT USA, LLC recalls The Optima Coil System consists of an implantable embolizati

Recalled Product

The Optima Coil System consists of an implantable embolization coil comprised of a platinum-tungsten alloy attached to a proximal stainless steel hypo-tube and distal body coil delivery pusher with a radiopaque distal positioning marker and proximal fluorosafe marker. The Optima Coil delivery pusher is 185cm in length. it is designed for use with the XCEL Detachment Controller. The XCEL Detachment Controller comes pre-loaded with batteries and is a sterile, handheld, single-patient-use device which should not to be reused, re-sterilized, opened or tampered with. The Optima Coil System and XCEL Detachment Controller(s) are sold separately. The Optima Coil achieves detachment by an internal heater element, which is powered by the XCEL Detachment Controller.


Hazard / Issue

Due to Label/Pouch mix-up. Pouches incorrectly labeled. Incorrect shelf carton label or pouch label due to inadequate manufacturing line clearance. As a result of this labeling mix-up, there is the possibility of inadvertently selecting the incorrect coil size for the procedure potentially causing some harms like vessel damage, an inability to treat with said device and potential rupture of the aneurysm.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model/REF: OPTI0204CSF10 Lot Number: F210500134 UDI Code: 00818053026010 Model/REF: OPTI0407CSF10 Lot Number: F210500135 UDI Code: 00818053026096