medical Class II Updated 2022-03-02

Natus Neurology DBA Excel Tech., Ltd. (XLTEK) recalls IPELA FHD PTZ TILE KITS with Axis cameras used in combinatio

Recalled Product

IPELA FHD PTZ TILE KITS with Axis cameras used in combination with Natus Neuroworks/Sleepworks software


Hazard / Issue

Tile kits that allow one to install and situate a video camera within a patient environment are not fitting properly in some domestic institution ceiling tile grids that have a thinner style grid frame. The ceiling tile may suddenly fall from the ceiling and cause a safety risk to the patient or end user.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All serial numbers
View official government recall