medical MODERATE Updated 2024-12-18

Micro-X Ltd. recalls MICRO-X Rover Mobile X-ray System, # MXU-RV35

See all recalls from Micro-X Ltd. →

Recalled Product

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 9357123000037. Serial Numbers: 00334, 00348, 00349, 00351, 00361.¿

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for MICRO-X Rover Mobile X-ray →
View official government recall