medical Class II Updated 2019-12-18

Boston Scientific Corporation recalls Habib EUS RFA, 10mm electrode active tip, Material Number/UP

Recalled Product

Habib EUS RFA, 10mm electrode active tip, Material Number/UPN: 6500 - Product Usage: The Habib EUS RFA is indicated for coagulation and ablation of soft tissue when used in conjunction with compatible radiofrequency generator.


Hazard / Issue

During testing on the Habib EUS RFA device, a temperature increase was observed at the tip of the EUS-FNA needle. It was determined that capacitive 5 coupling presents a risk of energy transfer from the Habib EUS RFA active tip to the introducer device. This capacitive coupling could result in thermal injury to tissue outside of the targeted treatment area due to its use with an un-insulated EUS-FNA needle requiring little or no medical intervention.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Material Number/UPN: 6500 GTIN: 5060271140725 Batches: 17-C02 (Expiration: 9-Jul-2020), 17-C06 (Expiration: 19-Feb-2021), 17-C07 (Expiration: 19-Feb-2021), 17-C08 (Expiration: 20-Feb-2021)
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →