medical Class I Updated 2025-01-01

Nova Biomedical Corporation recalls StatStrip Glucose Hospital Meter System. Model/Catalog Numbe

Recalled Product

StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.


Hazard / Issue

A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model No 63685. UDI-DI (01)10385480636858. Software Versions: v0.0.13.10 to v0.0.13.44