medical Class II Updated 2019-01-02

Elekta, Inc. recalls MOSAIQ Oncology Information System Version 2.64. MOSAIQ

Recalled Product

MOSAIQ Oncology Information System Version 2.64. MOSAIQ is an oncology information system used to manage workflows for treatment planning and delivery


Hazard / Issue

If a Setup field (CT, kV, MV or MVCT) is created by COPYING a treatment field or another setup field, then the copied field will have the same isocenter as the parent and that isocenter may not be the correct isocenter for imaging when it is sent to the machine.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Software Build(s): 2.64 SP9 and higher; UDI (01)00858164002091(10)2.64.235
View official government recall