medical Class II Updated 2019-12-25

Teleflex Medical recalls Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+

Recalled Product

Arrow EZ-IO Intraosseous Vascular Access System 25mm Needle+Stabilizer Kit, Product Code 9001P-VC-005 - Product Usage: The Arrow EZ-IO Intraosseous Vascular Access System provides intraosseous access in the proximal humerus, proximal tibia, and distal tibia of adult and pediatric patients, and the distal femur in pediatric patients when vascular access is difficult to obtain in emergent, urgent, or medically necessary cases for up to 24 hours.


Hazard / Issue

The safety cap attached to needles within the needle sets may become dislodged exposing the needle and potentially causing the needle to protrude through the packaging. If this issue is not detected, the immediate risk of exposure to the affected devices is needle stick injury to the clinician or health care professional. In addition, a puncture of the packaging may compromise the sterility of the needle.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →