Icotec Ag recalls VADER pedicle system, torque wrench, Catalog Number 42-702
Recalled Product
VADER pedicle system, torque wrench, Catalog Number 42-702
Hazard / Issue
icotec has received notification of one revision that had to be carried out because nut screws had become loose. During the investigation of the device involved, it was detected that the torque limiting function of the wrench used to tighten the nut screws during implantation was not functioning correctly. Due to the malfunction, the nut screws may not have been tightened sufficiently.
Issued by
FDA
Affected States: AZ, MI, OH, CA, CO, TX, NY, NC, DE, PA