medical Class II Updated 2020-12-30

ICU Medical, Inc. recalls Spinning Spiros Closed Male Luer, Red Cap, REF: IB-CH2000SC;

Recalled Product

Spinning Spiros Closed Male Luer, Red Cap, REF: IB-CH2000SC; 5" (13 cm) Appx 1.5 ml, Add-On Set w/Spiros, REF: IB-CH3034; 5" (13 cm) Appx 1.5 ml, Bag Spike Adapter w/Spiros, Vented Cap, REF: IB-CH3034C


Hazard / Issue

Potential for spinning versions of male luers to exhibit leaks due to a molding defect. Fluid leakage may potentially cause delay of infusion, contamination of the fluid path, exposure to hazardous medications, blood loss, exposure to patient blood, or fluid path air-in-line.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lots: 4605119, 4726277, 4727007 ,
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →