medical Class II Updated 2022-03-09

Abbott Molecular, Inc. recalls Alinity m Integrated Reaction Units (IRU)

Recalled Product

Alinity m Integrated Reaction Units (IRU)


Hazard / Issue

There is potential for the Reaction Vessels (RVs) detaching from the RV cap during transport by the pipettor robot.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: List Number 09N26-010, Lot Numbers/UDI: 1) 766233 / (01)20884999047864(10)766233(11)210614(240)09N26-010 2) 770433 / (01)20884999047864(10)770433(11)210621(240)09N26-010 3) 771200 / (01)20884999047864(10)771200(11)210712(240)09N26-010 4) 771201 / (01)20884999047864(10)771201(11)210719(240)09N26-010 5) 773642 / (01)20884999047864(10)773642(11)210723(240)09N26-010