medical Class II Updated 2022-03-09

Microbiologics Inc recalls KWIK-STIK, Quality control kit for culture media, sold as:

Recalled Product

KWIK-STIK, Quality control kit for culture media, sold as: a. KWIK-STIK 2-pack Enterococcus faecalis derived from ATCC 51299 b. KWIK-STIK 6-pack Enterococcus faecalis derived from ATCC 51299


Hazard / Issue

This strain should be resistant to high levels of Gentamicin and be resistant to Gentamicin Synergy. A specific lot was found susceptible to Gentamicin Synergy during in-house testing.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: a. Catalog Number 0959P, UDI: 20845357023210, Lot Number (Expiration Date): 959-74-2 (10/08/2014), 959-74-3 (10/08/2014), 959-74-7 (10/08/2014), 959-75-4 (04/22/2015), 959-75-8 (04/22/2015), 959-75-9 (04/22/2015), 959-75-14 (04/22/2015), 959-76-1 (07/08/2015), 959-77-3 (10/21/2015); b. Catalog Number 0959K, UDI: 30845357023224, Lot Number (Expiration Date): 959-74-4 (10/08/2014), 959-74-6 (10/08/2014), 959-74-8 (10/08/2014), 959-75-1 (04/22/2015), 959-75-8 (04/22/2015), 959-75-12 (04/22/2015), and 959-77-1 (10/21/2015)