medical Class II Updated 2020-12-30

Beckman Coulter Inc. recalls MicroScan Pos MIC Panel Type 34 - Product Usage: For the det

Recalled Product

MicroScan Pos MIC Panel Type 34 - Product Usage: For the determination of antimicrobial agent susceptibility and/or identification to the species level of rapidly growing aerobic and facultative gram-positive cocci, some fastidious aerobic cocci and listeria monocytogenes.


Hazard / Issue

Due to a customer report and internal investigation, it was confirmed that a portion of the affected product lot failed quality control (QC) for ATCC E. Faecalis 29212 versus daptomycin and it is a possibility to obtain false elevated MIC for clinical isolates with daptomycin even when quality control results are within specification.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalog Number: B1017-216 UDI Code: (01)15099590657550(17)210610(11)200610(10)2021-06-10 Lot Number: 2021-06-10
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →