Philips North America Llc recalls Philips Allura Xper, UNIQ(Allura Xper system-China) Interve
Recalled Product
Philips Allura Xper, UNIQ(Allura Xper system-China) Interventional Fluoroscopic X-ray System products with a Laird chiller WITHOUT an extended drip tray: Model Number: 722003, 722005, 722006, 722008, 722010, 722011, 722012, 722013, 722014, 722015, 722019, 722020, 722022, 722023, 722024, 722025, 722026, 722027, 722029, 722033, 722034, 722036, 722038, 722039. Allura Xper F010 {C+F). 722003 Allura Xper F010/10 722005 Allura Xper F020 722006 Allura Xper F020/10 and F020/20 722008 Allura Xper F010 722010 Altura Xper F010/10 722011 Allura Xper F020 722012 Allura Xper FD20 biplane 722013 Allura Xper FDlO OR Table 722014 Allura Xper FD20 OR Table 722015 Allura Xper FDl0/10 OR Table 722019 Allura Xper FD20 Biplane OR Table 722020 Allura Xper FD10 OR Table 722022 Allura Xper FD20 OR Table 722023 Allura Xper FD10/10 OR Table 722024 Altura Xper FD20 Biplane OR Table 722025 Allura Xper FD10 722026 Allura Xper FD10/10 722027 Altura Xper FD20/10 biplane 722029 Allura Xper FD10 OR Table 722033 Allura Xper FD10/10 OR Table 722034 Allura Xper FD20/10 OR Table 722036 Allure Xper F020/20 722038 Altura Xper FD20/20 biplane OR Table 722039
Hazard / Issue
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
Issued by
FDA
Were you affected by this recall?
If you were harmed by a recalled product, you may have legal options. Consider speaking with a consumer protection lawyer in your state.
Looking for safe alternatives?
Browse certified-safe products in this category on Amazon →
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.
CAN-SPAM compliant · No spam · Unsubscribe anytime