medical Class II Updated 2022-03-16

Philips North America Llc recalls Philips Allura CV20 Interventional Fluoroscopic X-ray System

Recalled Product

Philips Allura CV20 Interventional Fluoroscopic X-ray System products with a Laird chiller without an extended drip tray: Model Number: 722031


Hazard / Issue

Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: S/N: OUS 109
View official government recall