medical Class II Updated 2022-03-16

Heartware, Inc. recalls HeartWare Ventricular Assist Device (HVAD) System

Recalled Product

HeartWare Ventricular Assist Device (HVAD) System


Hazard / Issue

Updated cleaning instructions are being provided for the Controller AC Adapter, DC Adapter, and Battery to provide clarity to avoid cleaning the power source connector pins, as this could remove the lubricant that is applied to the pins as a mitigation for power switching.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots/serials for the following product model numbers: 1440, 1400, 1401, 1401AU, 1401BR, 1401CA, 1401CH, 1401DE, 1401GB, 1401IL, 1401IN, 1401IT, 1401JP, 1401US, 1403US, 1407AR, 1407AU, 1407BR, 1407CA, 1407CA-CLIN, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1425AU, 1425CA, 1425DE, 1425GB, 1425IL, 1425IT, 1425US, 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, 1430US, 1440, 1650, 1650, 1650CA-CLIN, 1650DE
View official government recall