Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ;
Recalled Product
epoc NXS Host; Siemens Material Number (SMN): 11413879(IN) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.
Hazard / Issue
Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.
Issued by
FDA
Were you affected by this recall?
If you were harmed by a recalled product, you may have legal options. Consider speaking with a consumer protection lawyer in your state.
Looking for safe alternatives?
Browse certified-safe products in this category on Amazon →
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.