medical Class II Updated 2024-01-24

Smiths Medical Asd Inc recalls Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX

Recalled Product

Medex 20IN (50.8CM) INJ LINE ROTATING ADAPT, List Number MX682R


Hazard / Issue

There is a manufacturing defect affecting specific rotator lots. Within this population, the inner diameter of the devices O-ring may be oversized affecting seal integrity. An inadequate seal may lead to a leak during infusion of medication which may potentially lead to under infusion of medication.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number: 4398433
View official government recall