medical Class I Updated 2019-02-06

WEST PHARMA. SERVICES IL, LTD recalls Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only

Recalled Product

Vial2BAG DC 13mm, REF 6070111, Single use only, Rx Only


Hazard / Issue

Adverse events have been reported with the use of oxytocin in conjunction with the product in pregnant women. There is a possibility that the device may cause inadequate admixing of drug with the solution in the IV bag. This could lead to concentrated drug pooling in the device and bolus injections.

Issued by

FDA

Affected States: MO
Lot/Code Info: UDI: (01)07290108240061(17)211031(10)A839(90)3609812 5(91)6070111 Lot Numbers: 7153, 7154, 7634, 7688, 7752, 7909, 7955, 8152, 8153, 8154, 8232, 8233, 8234, 8235, 9028, 9175, 9029, 9176, 9177, 9178, 9367, 9368, 9369, 9370, 9654, 9656, 9655, 9918, 9710, 9711, 9712, 9713, 9839, 9840, 9983, 9984, 9992, 9993, 9994, 9995, A123, A124, A790, A805, A806, A791, A792, A793, A794, A795, A796, A797, A807, A808, A809, A810, A834, A835, A836, A837, A838 & A839
View official government recall