medical Class II Updated 2025-01-01

Siemens Healthcare Diagnostics, Inc. recalls Magnesium Flex reagent cartridge, 360 tests, Catalog Number:

Recalled Product

Magnesium Flex reagent cartridge, 360 tests, Catalog Number: K3057; Siemens Material Number (SMN): 10445158;


Hazard / Issue

Siemens Healthcare Diagnostics is recalling two lots of their Dimension Vista¿ Total Magnesium (MG) Flex Reagent Cartridges (lots 24057BB and 24064BC) due to the potential for falsely low measurements of magnesium in plasma. Because the error is intermittent, existing quality control procedures may not recognize that a patients results are incorrect, and the results may be reported out

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Siemens Material Number (SMN): 10445158; Catalog Number: K3057; UDI-DI: 00842768015694; Lot numbers: 24057BB, 24064BC.
View official government recall