medical Class I Updated 2024-02-07

Datex-Ohmeda, Inc. recalls EVair 03 (Jun-air) air compressor, model numbers: 1609000 a

Recalled Product

EVair 03 (Jun-air) air compressor, model numbers: 1609000 and 1609002


Hazard / Issue

GE HealthCare has become aware of elevated levels of formaldehyde when the EVair or EVair 03 (Jun-Air) optional compressors are used with the CARESCAPE R860 or Engstr¿m Carestation/Pro ventilators, respectively. These elevated results were observed in preliminary testing that was conducted at an elevated room temperature of 40oC (104oF), at the lowest possible flow condition of 2 L/min (worstcase, minimum bias flow with no additional ventilation), and all of the gas being supplied from the compressor (i.e., FiO2 of 21% / no supplemental oxygen). GE HealthCare is continuing to evaluate the root cause for these elevated levels of formaldehyde. While these test conditions are not representative of typical clinical use conditions,GE HealthCare is taking this action to further reduce the potential for patient exposure.

Issued by

FDA

Distribution: Worldwide
Lot/Code Info: UDI/DI ********, Serial Numbers: CBCS01269, CBCT02328, CBCT03050, CBCT02306, CBCT02301, CBCT02303, CBCT02302, CBCT02088, CBCT02309, CBCT02304, CBCU00717, CBCU00559