medical Class II Updated 2021-01-13

Boston Scientific Corporation recalls Hot AXIOS Stent and Electrocautery - Enhanced Delivery Syste

Recalled Product

Hot AXIOS Stent and Electrocautery - Enhanced Delivery System 10mm x10mm (OUS) UPN: M00553540 - Product Usage: The stent is intended for implantation up to 60 days and should be removed upon confirmation of pseudocyst or walled-off necrosis resolution.


Hazard / Issue

The rotating luer fitting at the distal end of the delivery system handle may become detached from the nose of the delivery during device preparation, stent delivery or use, resulting in delay or procedure or risk of peritonitis developing from leakage of enteric, pseudocyst, or biliary contents into the peritoneal cavity

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number: 26130517 Exp. Date Range: 10/5/2022 GTIN: 08714729904564