Abbott Gmbh & Co. KG recalls Alinity i Processing Module, (a) Part Number 03R65-01 (b)

Recalled Product

Alinity i Processing Module, (a) Part Number 03R65-01 (b) Part Number A-30108552-01 (c) Part Number A-30109694-01 Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples.


Hazard / Issue

Abbott has identified an issue with the Alinity i Gear Pump Assembly (part number A 30108552 01) resulting in foaming/bubbling out of the bottle reservoir for Concentrated Wash Buffer and an unexpected amount of dried residue of buffer. The foaming/bubbling is a result of an improper calibration of the pump s operating speed.

Issued by

FDA

Distribution: Worldwide Distribution
Lot/Code Info: (a) Serial Numbers: Ai01001 - Ai01455, Ai01457, Ai01459 (b) all Serial Numbers (c) all Serial Numbers